| Photometry | | Top irradiation / Top reading fluorometry |
| Measuring Direction | | Vertical or Horizontal |
| Ex light focus | | Adjustable altitude(2-step) |
| Interference Filter | | Excitation (Ex) : 365, 490 nm (Up to five additional filters) Emission (Em) : 450, 530 nm (Up to five additional filters) |
| Wavelength range | | Excitation (Ex) : 260 - 710 nm / Emission (Em) : 300 - 750 nm |
| Temperature Control Function | | Room ambient, 37ºC, 40ºC and 45ºC (Accuracy ±0.5ºC) |
| Mixing Function | | Round/Linear, 3-step intensity |
| Mixing Time Setting | | 5, 10 and 20 seconds |
| Light Source | | Xenon Lamp 150W |
| Detector | | Photomultiplier |
| Power Consumption | | AC100V (50/60Hz) / 420VA |
| Dimension and Mass | | 400(W)×450(D)×315(H)mm Approx. 28kg (Standard model) |
|